Ionic (Electrochemical)MIGRATION EVALUATION SYSTEM AMI SERIES
  • Detects changes in insulation resistance with high precision
    Detects decline of insulation resistance by continuous measurement while applying voltage under high-temperature, high- humidity conditions.
  • Excels in detection of leak current
    The leak-touch detection function detects the Occurrence of ion migration in microsecond intervals. It immediately shuts off stress voltage to the channel when leak-touch is detected. You may choose whether to continue testing after detection (optional).
  • Precise insulation resistance
    measurement Over a wide range of 2105 to 11013 at end of the 3m measurement Cable (for AMI-025-S).
Resistance measuring range
(Direct current measurement range)
2105 to 11013(at 100V charge)
Minimum measuring level : 2103(at DC 1V charge)
(10pA to 500A)
Current leak detection range 1 to 500A
Measuring applied voltage 1V to 100V DC
(0.1V step)
Stress charge voltage 1 to 100V DC
(0.1V step, set secretory from measurement voltage)
Continuous testing time Maximum 9.999hours
Measuring time 60 sec/25ch + charge time
(set by user)
Current leak detection interval Adjustable (except when measuring resistance)
Measurement speed : sec order
(regardless of number of channels)
Number of measurement channels 25 channels (Max150)